RIGOL 
© Copyright Rigol Technologies, Inc. 2007. 
  User Manual for DM3000 Series 
6-4 
Chapter 6 : Appendix 
Appendix A: Specifications 
 
DM3000 Series Digital Multimeters 
Range: 
Or  
Performance Parameters
Resolution, 
Or  
AC Voltage Frequency:
Accuracy: 
1 Year
±(% of reading + 
% of range) 
Input Current, 
Or Current Source:
 
DC Voltage 
Input Current 
200.000,0 mV    100nV  0.0050 + 0.0017 
10MΩ 或>10GΩ 
2.000,000 V  1μV    0.0040 + 0.0004 
10MΩ 或>10GΩ 
20.000,00 V  10μV    0.0035 + 0.0003 
10MΩ 或>10GΩ 
200.000,0 V  100μV    0.0045 + 0.0003  10MΩ 
1000.000 V  1mV  0.0045 + 0.005  10MΩ 
DC Current 
Input Current 
2.000,00 mA  10nA  0.005 + 0.005    50Ω 
20.000,0 mA    100nA  0.005 + 0.002    50Ω 
200.000 mA    1μA  0.03 + 0.003    1Ω 
1.000,00 A    10μA  0.03 + 0.006    1Ω 
10.000,0 A    100μA  0.05 + 0.01    0.01Ω 
AC Voltage(RMS)    
200.000 mV  3Hz-5Hz  0.10 + 0.015     1MΩ 
   5Hz-10Hz  0.06 + 0.015     1MΩ 
   10Hz-20kHz  0.04 + 0.015     1MΩ 
   20kHz-50kHz  0.10 + 0.025     1MΩ 
   50kHz-100kHz  0.55 + 0.04     1MΩ 
   100kHz-300kHz  1.20 + 0.25     1MΩ 
Range from 2.000,00V to
750.000  
3Hz-5Hz
  1.00 + 0.01     1MΩ