RIGOL 
2-16 © Copyright RIGOL Technologies, Inc. 2007.       
  User Manual for DM3000 Series 
2.  Limit Measurement 
The Limit test function enables you to perform pass/fail testing to upper and lower 
limits that you specify. You can set the upper and lower limits to any value between 
0 and ±120% of the highest range, for the present function. The upper limit you 
select must be a more positive number than the lower limit.   
 
 
Press     Æ    Limit , enter the menu shown below: 
 
Figure 2-15 
 
Table 2-12 Limit Measurement Menu Function Explanation 
 
The parameters value scope of Limit function: 
(1). The limit value scope is 0%~±120% of the current measurement range. 
(2). The upper limit value should be always bigger than the lower limit value. 
 
Function Menu  Settings  Explanation 
High    Set the desired Upper limit. 
Low    Set the desired Lower limit. 
 
  Save all changes, back to a higher level menu.