Time Measurements 
The DS1000B series oscilloscopes provide 12 kinds timing parameters 
auto-measurements; Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay 
AėB
, Delay AėB , +Duty and –Duty, Phase AėB ,Phase AėB . 
Figure 2-124 shows a pulse with some of the time measurement points. 
 
 
Figure 2-124 The time parameters 
Rise Time: Time that the rising edge of the first pulse in the waveform takes to rise 
from 10% to 90% of its amplitude. 
Fall Time: Time that the falling edge of the first pulse in the waveform takes to fall 
from 90% to 10% of its amplitude. 
+Width: The width of the first positive pulse in 50% amplitude points. 
-Width: The width of the first negative pulse in the 50% amplitude points. 
Delay AėB
: The time delays between the channel A and channel B at the rising 
edge. 
Delay AėB
: The time delays between the channel A and channel B at the falling 
edge. 
Phase A
B : The phase between the channel A and channel B at the rising edge. 
Phase=(delay/source A period)h360e 
Phase A
B : The phase between the channel A and channel B at the falling edge. 
+Duty: +Duty Cycle, defined as +Width/Period 
-Duty: -Duty Cycle, defined as -Width/Period.