Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800/E Series   
3.  Save Preset 
Save the current instrument setting as user-defined setting into the internal 
non-volatile memory. You can store and name up to 6 system states 
(correspond to “User1” to “User6” in the preset type).  
 
When one of “User1” to “User6” is selected in Preset Type, press Save 
Preset and the instrument automatically opens the filename input interface 
(refer to “
To Input Filename” to store the setting).  
 
  Note: This menu is grayed out and disabled when “Factory” is selected in 
Preset Type. 
 
 
Calibrate 
 
1.  Cal Now 
Press this key and the analyzer will use the internal calibration source to 
perform a self-calibration immediately. Self-calibration would last about 5 s 
and “Calibrating” is displayed in the user interface status bar during the 
calibration. 
 
2.  Self-Cal 
When auto calibration is enabled, the analyzer executes self-calibration 
regularly. Within half an hour after power-on, the analyzer executes a 
self-calibration every 10 minutes and per hour after being powered on for 
more than half an hour. 
 
3.  Acc Cal 
Press this key and the analyzer executes comprehensive calibration which 
includes the amplitude calibration when the preamplifier is turned on/off and 
the spurious signal calibration. “Calibrating” is displayed in the user interface 
status bar during the calibration.