Chapter 5 Specifications  RIGOL 
User’s Guide for DSA800/E Series   
Chapter 5  Specifications 
 
Specifications are valid under the following conditions: the instrument is within the 
calibration period, is stored for at least two hours at 0℃ to 50℃ temperature and 
is warmed up for 40 minutes. Unless otherwise noted, the specifications in the 
manual include the measurement uncertainty. 
 
Typical (typ.): characteristic performance, which 80 percent of the measurement 
results will meet at room temperature (approximately 25℃). This data is not 
warranted and does not include the measurement uncertainty. 
 
Nominal (nom.): the expected mean or average performance or a designed 
attribute (such as the 50Ω connector). This data is not warranted and is measured 
at room temperature (approximately 25℃). 
 
Measured (meas.): an attribute measured during the design phase which can be 
compared to the expected performance, such as the amplitude drift variation with 
time. This data is not warranted and is measured at room temperature 
(approximately 25℃). 
 
Note: All charts in this manual are the measurement results of multiple instruments at room 
temperature unless otherwise noted.