60
Serial Interface
CAL n Set calibration mode.
n = 0 Calibration OFF
n = 1 Calibration ON
CAL? Get calibration mode status
Output data format
p<EOT><CR><LF>
p = 0 Calibration OFF
p = 1 Calibration ON
BAT? Get battery capacity status
Output corresponds to status of battery capacity indicator
on display.
Output data format
p<EOT><CR><LF>
p = 0 Flashing
p = 1 1 segment lit
p = 2 2 segments lit
p = 3 3 segments lit
DOD? Get measurement data
Returns a value corresponding to the display indication.
Output data format
p1 p2 p3 p4 E ± p5, p6 <EOT><CR><LF>
p1 to p4 Effective 4 digits of measurement value
p5 = -6 to +1
p6 = O (overload occurred)
p6 = _ (no overload occurred) (_ is a space)
* Explanation of p5
p5 stands for 10
p5
.
For example, when measurement value is 10000, p5
will be 1, because 1000 × 10
1