20
Measurement
Power-On
When the unit is turned on, the same settings as used before the unit was turned
off are reestablished (resume function), and measurement starts. The level range,
high-pass filter setting, low-pass filter setting, and display characteristics are
memorized for each measurement mode.
Settings memorized by Settings reset at regular
resume function vibration measurement start*
Measurement mode CAL status
Level range Maximum value hold
Display characteristics Peak hold
High-pass filter Backlight status
Low-pass filter Sensitivity setting status
Sensitivity MENU setting status
MENU0 to MENU2 item settings Comparator status
MENU4 to MENU8 item settings Remote status
* These measurement conditions are not memorized by the resume
function. They will be reset to the default vibration measurement con-
dition when the unit is turned on.
Note
Settings made via the serial interface are not memo-
rized for the resume function.