- 63 - GD-70D
<Setpoint Indicator "1-2">
Use this menu item to check important setpoints. Setpoints cannot be changed in User Mode.
Setpoint Display
PW
A1
A2
F
LCD
From Detection Mode, press the
MODE key for three seconds.
0.0ppm
SIH4
1-1.ZERO
Press the key.
1- 1
ZERO
MAINTENANCE
1-2.CONFIRM
Press the SET key.
1- 2
CONFIRM
MAINTENANCE
First Alarm Setpoint
5.0ppm
AL 1
MAINTENANCE
Second Alarm Setpoint
10.0ppm
AL 2
MAINTENANCE
Alarm Delay Time (seconds)
2
ALM DLY
MAINTENANCE
Zero Suppression Value
0.9ppm
SUPPRESS
MAINTENANCE
Zero Follower ON/OFF
(If ESU or SSU sensor installed)
ON
ZERO F
MAINTENANCE