52 • Calibrating A GX-2003 With the Data Logging Software
To calibrate a GX-2003, perform the following steps:
1. Launch the GX-2003 Data Logging Software.
2. .Connect a GX-2003 to the software. See Downloading
Data from the GX-2003 on page 23, instruction 1 through
instruction 3, for instructions to connect an instrument to the
software.
3. Once the connection is made, the Complete Download,
Instrument Information, and Power Off download
commands will be selectable. Click the Instrument
Information download command to retrieve the instrument
information from the GX-2003. If you wish to download data
before calibrating, click Complete Download instead of
Instrument Information.
4. Click the Set button to display the Set window.
5. Click the Calibration button in the GX-2003 Status frame.
The Calibration frame will replace the GX-2003 Status frame
and the instruments pump will begin to operate.
Figure 36: Set Window
Click to begin calibration of a GX-2003