GX-2003 Operator’s Manual Calibration Mode • 39
WARNING: If you are using a calibration kit that includes a gas bag
and a fixed flow regulator or dispensing valve, do not
apply gas directly to the GX-2003 with the regulator or
valve or damage to the pump will result. See “Appendix
A: Calibrating with a Sample Bag” on page 86 for
instructions to properly use a gas bag kit.
To calibrate the %LEL, Oxygen, CO, and H2S sensors at the same
time, automatically, with no need for a zero-oxygen source, you can
use the auto calibration feature with a 4-gas cylinder. If the H2S
channel is not active, then a 3-gas cylinder may be used for auto-
calibration. This section includes instructions for auto calibration with
a demand-flow regulator and a 4-gas cylinder. This section also
includes instructions for calibrating one channel at a time using single
calibration.
Preparing for Calibration
Whether you will calibrate the Model GX-2003 using auto calibration
or single calibration, you will need to set the fresh air readings first
before setting the span readings for each gas.
1. Find a fresh-air environment. This is an environment free of toxic
or combustible gasses and of normal oxygen content (20.9%).
2. Turn on the unit as described inTurning On the GX-2003, Normal
Mode Only” on page 20 using the sample hose and probe.
3. With the unit on, perform a demand zero by pressing and holding
the AIR button. The display prompts you to hold the AIR
button.
4. Continue to hold the AIR button until the display prompts you
to release the AIR button. The GX-2003 will set the fresh air
reading for all channels.
5. Install the demand flow regulator onto the calibration cylinder.
6. Connect the sample tubing to the demand flow regulator.
Calibrating the GX-2003
From Measuring Mode, enter Calibration Mode by pressing and
holding the (SHIFT) button, then pressing the DISPLAY(ADJ) button
and releasing both. If the unit prompts you for the password, enter it
by using the AIR and (SHIFT) buttons to select each password
number and then pressing and releasing POWER ENTER to enter it
and move on to the next one. The calibration menu displays with the
cursor next to AUTO CALIBRATION.

Table of Contents

Other manuals for RKI Instruments GX-2003

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-2003 and is the answer not in the manual?

RKI Instruments GX-2003 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-2003
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals