66 • Chapter 5: User Mode and Calibration GX-3R Operator’s Manual
AUTO CAL
ON (factory setting): If a bump test fails, the unit automatically begins a calibration.
OFF: If a bump test fails, the unit does not automatically begin a calibration.
1 . After entering the BUMP SET\SETTINGS menu, press AIR to scroll to AUTO CAL.
2 . Press and release POWER MODE. The current setting flashes.
3 . Use AIR to display the desired setting.
4 . Press and release POWER MODE to save the setting and return to the AUTO CAL
menu item.
5 . See “Exiting the SETTINGS Menu” on page 66 to return to the BUMP SET menu.
6 . See “Exiting the BUMP SET Menu” on page 67 to return to User Mode.
Exiting the SETTINGS Menu
1 . While in the BUMP SET\SETTINGS menu, press AIR to scroll to ESCAPE.
2 . Press and release POWER MODE. The instrument returns to the SETTINGS menu item
of the BUMP SET menu.
BUMP.RMDR
ON: The GX-3R gives an indication at start up if it is due for bump testing. The type of
indication depends on the BUMP.EXPD setting (see pg.67).
OFF (factory setting): The GX-3R does not give an indication at start up if it is due for bump
testing.
1 . After entering the BUMP SET menu, press AIR to scroll to BUMP.RMDR.
2 . Press and release POWER MODE. The current setting flashes.
3 . Use AIR to display the desired setting.
4 . Press and release POWER MODE to save the setting and return to the BUMP.RMDR
menu item.
5 . See “Exiting the BUMP SET Menu” on page 67 to return to User Mode.