GX-3R Operator’s Manual Chapter 5: User Mode and Calibration 67
BUMP INT
This setting defines the amount of time between bump tests. The time can be set in 1 day
increments. The minimum setting is 0 days and the maximum setting is 30 days (factory
setting).
1 . After entering the BUMP SET menu, press AIR to scroll to BUMP INT.
2 . Press and release POWER MODE. The current setting flashes.
3 . Use AIR to display the desired setting.
4 . Press and release POWER MODE to save the setting and return to the BUMP INT menu
item.
5 . See “Exiting the BUMP SET Menu” on page 67 to return to User Mode.
BUMP.EXPD
This item defines what indication is given during start up when a bump test is due and
BUMP.RMDR is set to ON.
CONFIRM
(factory setting): The GX-3R gives an indication at start up if a bump test is past
due. Press and release AIR to continue without bump testing or press and release POWER
MODE to perform a bump test.
CANT USE
: The GX-3R gives an indication at start up if a bump test is past due. Press and
release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no
effect. A successful bump test must be performed in order to use the instrument.
NONE: The GX-3R gives an indication at startup if a bump test is past due. If desired, press
POWER MODE to perform a bump test but it is not necessary to acknowledge the bump test
due indication. The warm-up sequence will continue on its own.
1 . After entering the BUMP SET menu, press AIR to scroll to BUMP.EXPD.
2 . Press and release POWER MODE. The current setting flashes.
3 . Use AIR to display the desired setting.
4 . Press and release POWER MODE to save the setting and return to the BUMP.EXPD
menu item.
5 . See “Exiting the BUMP SET Menu” on page 67 to return to User Mode.
Exiting the BUMP SET Menu
1 . While in the BUMP SET menu, press AIR to scroll to ESCAPE.
2 . Press and release POWER MODE. The instrument returns to the BUMP SET menu item
in User Mode.
3 . See “Entering Measuring Mode (START)” on page 78 to enter Measuring Mode.
BUMP INT
M 7:49
BUMP.EXPD
M 7:49