FSE CALCulate Subsystem
1065.6016.12 6.59 E-16
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:PHOLd ON | OFF
This command switches on or off the peak-hold function.
Example: ":CALC:MARK:FUNC:SUMM:PHOL ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A-Z, VA
The peak-hold function is reset by switching off and on, again.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:AVERage ON | OFF
This command switches the calculation of the average value on or off.
Example: ":CALC:MARK:FUNC:SUMM:AVER ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A-Z, VA
The calculation of the average is reset by switching off and on, again.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:AOFF
This command switches off all measuring functions.
Example: ":CALC:MARK:FUNC:SUMM:AOFF"
Features: *RST value: _
SCPI: device-specific
Modes: A-Z, VA
This command is an "event"
and therefore has no *RST value assigned and no query.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:CENTer
This command sets the center frequency to that of the current marker.
Example: ":CALC:MARK:FUNC:CENT"
Features: *RST value: -
SCPI: device-specific
Mode: A-F
This command is an "event" and therefore has no *RST value assigned and no query.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:CSTep
This command sets the step width of the center frequency to the x-value of the current marker.
Example: ":CALC:MARK:FUNC:CST"
Features: *RST value: -
SCPI: device-specific
Mode: A-F
This command is an "event" and therefore has no *RST value assigned and no query.