FSE SENSe Subsystem
1065.6016.12 6.203 E-16
SENSe:POWer Subsystem
This subsystem controls the setting of the instrument’s power measurements.
COMMAND PARAMETER UNIT COMMENT
[SENSe<1|2>]
:POWer
:ACHannel
:SPACing
[:UPPer]
:ACHannel
:ALTernate<1|2>
:ACPairs
:BANDwidth
[:CHANnel]
:ACHannel
:ALTernate<1|2>
:BWIDth
[:CHANnel]
:ACHannel
:ALTernate<1|2>
:MODE
:REFerence
:AUTO
:PRESet
:BANDwidth
:BWIDth
<numeric_value>
<numeric_value>
<numeric_value>
1 | 2 | 3
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
ABSolute | RELative
ONCE
ACPower | CPOWer | OBANdwidth |
OBWidth | CN | CN0 | ADJust
<numeric_value>
<numeric_value>
HZ
HZ
HZ
HZ
HZ
HZ
HZ
HZ
HZ
PCT
PCT
no query
:[SENSe<1|2>:]POWer:ACHannel:SPACing[:UPPer] 0 Hz to 1000 MHz
This command defines the channel spacing of adjacent channel to carrier.
Example: ":POW:ACH:SPAC 28kHz"
Features: *RST value: 24 kHz
SCPI: conforming
Mode: A-F
:[SENSe<1|2>:]POWer:ACHannel:SPACing:ACHannel 0 Hz to 1000 MHz
This command defines the channel spacing of adjacent channel to carrier. This command has the
same effect as POW:ACH:SPAC.
Example: ":POW:ACH:SPAC:ACH 338kHz"
Features: *RST value: 24 kHz
SCPI: device-specific
Mode: A-F
:[SENSe<1|2>:]POWer:ACHannel:SPACing:ALTernate<1|2> 0 Hz to 1000 MHz
This command defines the spacing of the first (ALTernate1) or the second alternate adjacent
channel (ALTernate2) relative to the carrier signal.
Example: ":POW:ACH:SPAC:ALT1 99kHz"
Features: *RST value: 24 kHz
SCPI: device-specific
Mode: A-F