SENSe Subsystem FSE
1065.6016.12 6.202 E-16
:[SENSe<1|2>:]MSUMmary:REFerence <numeric_value>
This command selects the reference modulation.
Parameter: <numeric_value> := 0.001PCT to 1000PCT for AM
0.1HZ to 10MHZ for FM
0.0001RAD to 1000RAD for PM
Example: ":MSUM:REF 50PCT"
Features: *RST value: 100PCTfor AM
100KHZfor FM
10RAD for PM
SCPI: device-specific
Mode: VA-A
:[SENSe<1|2>:]MSUMmary:REFerence:AUTO ONCE
This command sets the current absolute measured values of the main modulation signal as
reference values for the relative indication.
Example: ":MSUM:REF:AUTO ONCE"
Features: *RST value: --
SCPI: device-specific
Mode: VA-A
This command is an event and has therefore no query and no *RST value assigned.
:[SENSe<1|2>:]MSUMmary:MTIMe 0.1 s | 1 s
This command selects the measurement time for the summary markers.
Example: ":MSUM:MTIM 100US"
Features: *RST value: 0.1S
SCPI: device-specific
Mode: VA-A