Status Reporting System FSE
1065.6016.12 5.34 E-16
Resetting Values of the Status Reporting System
Table 5-12 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST and SYSTem:PRESet influences the functional
instrument settings. In particular, DCL does not change the instrument settings.
Table 5-12 Resetting instrument functions
Event Switching on supply
voltage DCL,SDC
Power-On-Status-
Clear
(Device Clear,
Selected Device
Clear)
*RST or
SYSTem:PRESet
STATus:PRESet *CLS
Effect 0 1
Clear STB,ESR
yes

yes
Clear SRE,ESE
yes

Clear PPE
yes

Clear EVENTt parts of the
registers
yes

yes
Clear Enable parts of all
OPERation and
QUEStionable registers,
Fill Enable parts of all
other registers with "1".
yes

yes
Fill PTRansition parts with
"1" ,
Clear NTRansition parts
yes

yes
Clear error queue yes yes

yes
Clear output buffer yes yes yes 1) 1) 1)
Clear command
processing and input
buffer
yes yes yes

1) Every command being the first in a command line, i.e., immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer.