FSE SENSe Subsystem
1065.6016.12 6.167 E-16
SENSe:BANDwidth Subsystem
This subsystem controls the setting of the instruments filter bandwidths. Both groups of commands
(BANDwidth and BWIDth) perform the same functions.
COMMAND PARAMETERS UNIT COMMENT
[SENSe<1|2>]
:BANDwidth | :BWIDth
[:RESolution]
:AUTO
:MODE
:FFT
:RATio
:VIDeo
:AUTO
:RATio
:DEMod
:PLL
<numeric_value>
<Boolean>
ANALog|DIGital
<Boolean>
<numeric_value>
<numeric_value>
<Boolean>
<numeric_value> | SINe | PULSe |
NOISe
<numeric_value>
AUTO | HIGH | MEDium | LOW
HZ
--
--
--
--
HZ
--
--
HZ
option FFT Filter
Vector Signal Analysis
:[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution] <numeric_value>
This command defines the analyzers resolution bandwidth.
Example: ":BAND 1MHz"
Features: *RST value: - (AUTO is set to ON)
SCPI: conforming
Modes: A, VA, BTS, MS
The values for the resolution bandwidth are rounded in 1 | 2 | 3 | 5 steps.
In the GSM BTS/MS ANALYZER mode with option FSE-K11/K10, the command is available for
POWER vs. TIME measurement. In this case, the parameters DEFault (bandwidth setting according
to GSM standard), 300KHZ and 1MHZ are permitted.