FSE Status Reporting System
1065.6016.12 5.27 E-16
STATus QUEStionable:LIMit Register
This register comprises information about the observance of limit lines. It can be queried with
commands STATus:QUEStionable:LIMit:CONDition? and STATus:QUEStionable:LIMit
[:EVENt]?.
Table 5-7 Meaning of bits in STATus:QUEStionable:LIMit register
Bit No. Meaning
0
LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1
LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2
LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3
LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4
LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5
LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6
LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7
LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 not used
9 not used
10 not used
11 not used
12 not used
13 not used
14 not used
15 This bit is always 0.