DISPlay Subsystem FSE
1065.6016.12 6.98 E-16
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:RPOSition 0 to 100PCT
This command defines the position of the reference value.
Example: ":DISP:TRAC:Y:RPOS 50PCT"
Features: *RST value: 100PCT (tracking generator)
50PCT (vector analyzer)
SCPI: conforming
Modes: A, VA
The numeric suffix in TRACe<1 to 4> is not significant. This command is only valid in conjunction
with option Tracking Generator or in vector analyzer mode.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:PDIVision <numeric_value>
This command defines the scaling of the Y-axis in the current unit.
Example: ":DISP:TRAC:Y:PDIV +1.20"
Features: *RST value: -
SCPI: conforming
Mode: VA
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y:SPACing LINear | LOGarithmic | PERCent
This command toggles between linear and logarithmic display.
Example: ":DISP:TRAC:Y:SPAC LIN"
Features: *RST value: LOGarithmic
SCPI: conforming
Mode: A
The linear display can be LIN/% (PERCent) or LIN/dB (LINear). The numeric suffix in TRACe<1 to 4>
is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:MODE WRITe | VIEW | AVERage | MAXHold | MINHold
This command defines the type of display and the evaluation of the traces.
Example: ":DISP:TRAC3:MODE MAXH"
Features: *RST value: WRITe for TRACe1, STATe OFF for TRACe2 to 4
SCPI: device-specific
Modes: A, VA, BTS, MS