FSE INITiate Subsystem
1065.6016.12 6.125 E-16
INITiate Subsystem
The INITiate subsystem checks the initialization of the trigger subsystem.
In the split-screen representation, a distinction is made between INITiate1 (screen A) and INITiate2
(screen B)
COMMAND PARAMETERS UNIT COMMENT
INITiate<1|2>
:CONTinuous
:CONMeas
[:IMMediate]
:DISPlay
<boolean>
--
--
<boolean>
--
--
--
--
no query
no query
:INITiate<1|2>:CONTinuous ON | OFF
This command determines if the trigger system is continuously initiated ("Free Run").
Example: ":INIT:CONT OFF"
Features: *RST value: ON
SCPI: conforming
Modes: A, VA, BTS, MS
Setting "INITiate:CONTinuous ON" corresponds to function SWEEP CONTinuous, ie the sweep
of the analyzer is cyclically repeated. The setting "INITiate:CONTinuous OFF" corresponds to
function SWEEP SINGLE.
:INITiate<1|2>:CONMeas
This command continues the sweep from the current sweep position.
Example: ":INIT:CONM"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is an event and therefore has no *RST value and no query.
Sweeps are stopped, for example, in the case of transducer sets between the different transducer
factors.
:INITiate<1|2>[:IMMediate]
The command initiates a new sweep or starts a single sweep.
Example: "INIT"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.