READ-Subsystem FSE
1065.6016.12 6.160 E-16
Example: ":READ:SPEC:MOD?"
Result: 0,890E6,915E6,-87.4,-108.0,ABS,FAILED,
1,893.2E6,893.2E6,-83.2,-108.0,ABS,FAILED,
2,895.7E6,895.7E6,-87.4,-108.0,ABS,FAILED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
The command ABORt aborts an ongoing measurement.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the modulation spectrum is selected. (see
:CONFigure:SPECtrum:MODulation).
:READ:SPECtrum:SWITching[:ALL]?
This command starts the measurement of the transient spectrum of the base station or mobile and
reads out the result.
Parameter: The result is read out as a list of partial ASCII result strings separated by ’,’ in the
format used for READ:SPECtrum:MODulation[:ALL]?.
Example: ":READ:SPEC:SWIT?"
Result: 0,833.4E6,833.4E6,37.4,-36.0,ABS,MARGIN,
1,834.0E6,834.0E6,-35.2,-36.0,ABS,FAILED,
2,834.6E6,834.6E6,-74.3,-75.0,REL,FAILED
0,835.0E6,835.0E6,-65,0,-60.0,REL,PASSED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
The command ABORt aborts an ongoing measurement.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the transient spectrum is selected. (see :CONFigure:SPECtrum:SWITCHing).