READ-Subsystem FSE
1065.6016.12 6.162 E-16
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
The command ABORt aborts an ongoing measurement.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the spurious emissions is selected. (see :CONFigure:SPURious).
:READ:SPURious:STEP?
This command starts the next single step for measuring the spurious emissions in the STEP mode
and reads out the results. The measurement is performed in the currently set frequency range.
If the command READ:SPURious:STEP? is sent again after the last single step is reached, the
measurement sequence is terminated, the result of the last single step is output again and
characterized by the value ’FINISHED’ indicating its status. Until the status value ’FINISHED’ is
returned, the value ’RUNNING’ is output when the total result of the limit check is queried with the
command :CALCulate<1|2>:LIMit<1 to 8>:SPURious?.
Afterwards, sending the command again causes a restart of the measurement.
Parameter: The measured result is read out as a list of partial result strings separated by ’,’
and in the same format as for the command READ:SPURious[:ALL]?.
The additional status value ’FINISHED’ marks the end of a measurement sequence.
Example: ":READ:SPUR:STEP?"
Result: First query: 0,890E6,915E6,-87.4,-108.0,ABS,FAILED
...
Second-last query:1,893.2E6,893.2E6,-83.2,-108.0,ABS,FAILED
Last query: 1,893.2E6,893.2E6,-83.2,-108.0,ABS,FINISHED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
The command ABORt aborts an ongoing measurement. If the command READ:SPURious:STEP? is
sent again, the instrument restarts with the first single step.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the spurious emissions is selected. (see :CONFigure:SPURious).