SENSe Subsystem FSE
1065.6016.12 6.164 E-16
:[SENSe<1|2>:]ADEMod:SQUelch:LEVel 30 to -150 dBm
This command defines the switching threshold for the squelch referred to the measured signal.
Example: ":ADEM:SQU -10DBM"
Features: *RST value: -40dBm
SCPI: device-specific
Mode: VA-A
:[SENSe<1|2>:]ADEMod:SBANd NORMal | INVerse
This command selects the side band for the demodulation.
Example: ":ADEM:SBAN INV"
Features: *RST value: NORMal
SCPI: device-specific
Mode: VA-A
:[SENSe<1|2>:]ADEMod:RTIMe ON | OFF
This command selects whether the demodulation performed in real time or in blocks.
Example: ":ADEM:RTIM ON"
Features: *RST value: ON
SCPI: device-specific
Mode: VA-A