FSE SOURce Subsystem
1065.6016.12 6.213 E-16
:SOURce<1|2>:FM:STATe ON | OFF
This command switches on or off the external frequency modulation of the tracking generator.
Example: ":SOUR:FM:STAT ON "
Features: *RST-value: OFF
SCPI: conforming
Modes: A, VA
External AM, external I/Q-modulation and frequency offset are switched off, if active. This command
is only valid in conjunction with option Tracking Generator.
:SOURce<1|2>:FREQuency:OFFSet -200MHz to 200MHz
This command defines a frequency offset of the tracking generators to the current instrument
frequency.
Example: ":SOUR:FREQ:OFFS "
Features: *RST-value: 0 Hz
SCPI: conforming
Modes: A, VA
External delta modulation is switched off, if active. This command is only valid in conjunction with
option Tracking Generator.
:SOURce<1|2>:POWer:ALC:SOURce INTernal | EXTernal
This command switches on or off the external level control (ALC).
Example: ":SOUR:POW:ALC:SOUR INT "
Features: *RST- value: INT
SCPI: conforming
Modes: A, VA
External AM and external delta modulation are switched off, if active. This command is only valid in
conjunction with option Tracking Generator.
:SOURce<1|2>:POWer[:LEVel][:IMMediate][:AMPLitude] -20dBm to 0dBm /
Option FSE-B12: -90dBm to 0dBm
This command defines the level of the tracking generator.
Example: ":SOUR:POW -20dBm"
Features: *RST- value: -20 dBm
SCPI: conforming
Modes: A, VA
This command is only valid in conjunction with option Tracking Generator.
:SOURce<1|2>:POWer[:LEVel][:IMMediate]:OFFSet -200dB to +200dB
This command defines a level offset for the tracking generator.
Example: ":SOUR:POW:OFFS -10dB"
Features: *RST- value: 0dB
SCPI: conforming
Modes: A, VA