EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 339

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Assignment - General Device Settings FSE
1065.6016.12 6.282 E-16
DATA SET
CLEAR
:MMEMory:CLEar:STATe 1,<file_name>
DATA SET
CLEAR ALL
:MMEMory:CLEar:ALL
RECALL
EDIT
NAME
:MMEMory:LOAD:STATe 1,<file_name>
EDIT
PATH
the path is included in the file name.
AUTO
RECALL
:MMEMory:LOAD:AUTO 1,<file_name>
SELECT ITEMS
TO RECALL
--
SELECT
ITEMS
:MMEMory:SELect[:ITEM]:GSETup ON | OFF
:MMEMory:SELect[:ITEM]:HWSettings ON | OFF
:MMEMory:SELect[:ITEM]:TRACe<1...4> ON | OFF
:MMEMory:SELect[:ITEM]:LINes[:ACTive] ON | OFF
:MMEMory:SELect[:ITEM]:LINes:ALL ON | OFF
:MMEMory:SELect[:ITEM]:CSETup ON | OFF
:MMEMory:SELect[:ITEM]:HCOPy ON | OFF
:MMEMory:SELect[:ITEM]:CDATa ON | OFF
:MMEMory:SELect[:ITEM]:MACRos ON | OFF
:MMEMory:SELect[:ITEM]:SCData ON | OFF
:MMEMory:SELect[:ITEM]:TRANsducer[:ACTive] ON | OFF
:MMEMory:SELect[:ITEM]:TRANsducer:ALL ON | OFF
:MMEMory:SELect[:ITEM]:CVL[:ACTive] ON | OFF
:MMEMory:SELect[:ITEM]:CVL:ALL ON | OFF
ENABLE
ALL ITEMS
:MMEMory:SELect[:ITEM]:ALL
DISABLE
ALL ITEMS
:MMEMory:SELect[:ITEM]:NONE
DEFAULT
CONFIG
:MMEMory:SELect[:ITEM]:DEFault
DATA SET
LIST
--
DATA SET
CLEAR
:MMEMory:CLEar:STATe 1,<file_name>
DATA SET
CLEAR ALL
:MMEMory:CLEar:ALL

Table of Contents