List of Error Messages FSE
1065.6016.12 9.6 E-1
Continuation: Execution Error
Error code Error text in the case of queue poll
Error explanation
-270
Macro error
Error on the execution of a macro.
-271
Macro syntax error
The macro definition contains a syntax error.
-272
Macro execution error
The macro definition contains an error.
-273
illegal macro label
An illegal macro label is defined in the *DMC command.
Example: The label is too long. The label is identical with the common command header or contains an
invalid header syntax.
-274
Macro parameter error
The macro definition improperly uses a macro parameter placeholder.
-275
Macro definition too long
The macro definition is too long.
-276
Macro recursion error
The command sequence defined by the macro is trapped in a program loop.
Example: The event that would allow the loop to be exited does not occur.
-277
Macro redefinition not allowed
The macro label defined in the *DMC command is already defined elsewhere.
-278
Macro header not found
The macro label in the *GMC? query is not yet defined.
-280
Program error
Error on the execution of a down-loaded program.
-281
Cannot create program
The program cannot be created.
-282
illegal program name
The name of the programm is illegal.
Example: The name relates to a non-existing program.
-283
illegal variable name
The inputted variable does not exist in the program.
-284
Program currently running
The desired operation is not possible while the program is running.
Example: A running program cannot be deleted.
-285
Program syntax error
The down-loaded program contains a syntax error.
-286
Program runtime error

Table of Contents