FSE CALCulate Subsystem
1065.6016.12 6.35 E-16
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:STATe ON | OFF
This command activates the limit check for the first/second alternate adjacent channel for adjacent
channel power measurements. Before, the limit check must be activated using CALC:LIM:ACP ON.
Examples: ":CALC:LIM:ACP:ALT2:STAT ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:RESult?
This command queries the result of the limit check for the first/second alternate adjacent channel for
adjacent channel power measurements.
Parameter: The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED and where the first (second) returned value
denotes the lower (upper) alternate adjacent channel.
Examples: ":CALC:LIM:ACP:ALT2:RES?"
Features: *RST value: --
SCPI: device-specific
Modes: A, VA
This command is a query and therefore not assigned a *RST value. If the power measurement of the
adjacent channel is switched off, the command triggers a query error.
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.