Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
123User Manual 1173.9286.02 ─ 15
Manual operation: See "Top / Range / Bottom" on page 52
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel:OFFSet <Offset>
This command defines the trace offset.
Parameters:
<Offset> Range: -200 to 200
*RST: 0dB
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10
Defines a trace offset of -10 dB.
Manual operation: See "Trace Offset" on page 53
FORMat:DEXPort:DSEParator
<Separator>
This command selects the decimal separator for data exported in ASCII format.
Parameters:
<Separator> COMMa
Uses a comma as decimal separator, e.g. 4,05.
POINt
Uses a point as decimal separator, e.g. 4.05.
*RST: *RST has no effect on the decimal separator.
Default is POINt.
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation: See "Trace Export" on page 55
FORMat:DEXPort:HEADer <State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Manual operation: See "Trace Export" on page 55
FORMat:DEXPort:TRACes <Selection>
This command selects the data to be included in a data export file (see MMEMory:
STORe<n>:TRACe on page 124).
Graphical Display of Phase Noise Results