FETCh:PNOise:SPURs?
This command queries the location and level of all spurs that have been detected.
Return values:
<Spurs> Returns two values (frequency and level) for each each spur that
has been detected.
Example:
FETC:PNO:SPUR?
would return, e.g.
1999.232666,-0.639974,6494.312500,-0.760579,
19992.324219,-0.639974
Usage: Query only
Manual operation: See "Spur List" on page 17
FETCh:PNOise:SPURs:DISCrete?
This command queries the Discrete Jitter result.
Return values:
<Jitter> Default unit: s
Example:
FETC:PNO:SPUR:DISC?
would return, e.g.
2.3e-08
Usage: Query only
Manual operation: See "Spur List" on page 17
FETCh:PNOise:SPURs:RANDom?
This command queries the Random Jitter result.
Return values:
<Jitter> Default unit: s
Example:
FETC:PNO:SPUR:RAND?
would return, e.g.
3.59e-09
Usage: Query only
Manual operation: See "Spur List" on page 17
8.11.5 Reading Out Measured Values
FETCh:PNOise:MEASured:FREQuency?.........................................................................136
FETCh:PNOise:MEASured:LEVel?..................................................................................136