Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
140User Manual 1173.9286.02 ─ 15
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation: See "Marker 1 ... Marker x" on page 65
See " Marker Type " on page 66
CALCulate<n>:MARKer<m>:TRACe <Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Trace> 1 to 6
Trace number the marker is assigned to.
Example:
CALC:MARK3:TRAC 2
Assigns marker 3 to trace 2.
Manual operation: See " Assigning the Marker to a Trace " on page 66
CALCulate<n>:MARKer<m>:X <Position>
This command moves a marker to a particular coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Using Markers