Parameters:
<OffsetFrequency> The minimum offset is 1 Hz. The maximum offset depends on
the hardware you are using.
*RST: 1 MHz
Example:
CALC:EVAL:USER2:STAR 1khz
Defines a start point of 1 kHz for the second user range.
Manual operation: See "User Range" on page 57
CALCulate<n>:EVALuation:USER<range>:STOP
<OffsetFrequency>
This command defines the end point of a custom residual noise calculation range.
Before you can use the command, you have to assign the user range to a trace with
CALCulate<n>:EVALuation:USER<range>:TRACe on page 128.
Suffix:
<n>
.
Window
Parameters:
<OffsetFrequency> The minimum offset is 1 Hz. The maximum offset depends on
the hardware you are using.
*RST: 1 MHz
Example:
CALC:EVAL:USER2:STOP 100kHz
Defines an end point of 100 kHz for the second user range.
Manual operation: See "User Range" on page 57
CALCulate<n>:EVALuation:USER<range>:TRACe <Trace>
This command selects the trace for a custom residual noise calculation range.
Suffix:
<n>
.
Window
<range> 1...3
Selects the user range.
Parameters:
<Trace> NONE
Turns a user range off.
TRACE1 ... TRACE6
Trace to assign the user range to.
Example:
CALC:EVAL:USER2:TRAC TRACE2
Assigns the second user range to trace 2.
Manual operation: See "User Range" on page 57