Analysis
R&S
®
FSW-K40
58User Manual 1173.9286.02 ─ 15
Remote command:
Selecting a trace:
CALCulate<n>:EVALuation:USER<range>:TRACe on page 128
Define start frequency of user range:
CALCulate<n>:EVALuation:USER<range>:STARt on page 127
Define stop frequency of user range:
CALCulate<n>:EVALuation:USER<range>:STOP on page 128
6.2.2 Configuring Spot Noise Measurements
On All Decade Edges....................................................................................................58
On User Defined Offsets / Offset Frequency................................................................ 58
On All Decade Edges
Turns the calculation of spot noise on all 10
x
offset frequencies on and off.
Remote command:
Turn on and off spot noise calculation on 10
x
offset frequencies:
CALCulate<n>:SNOise:DECades[:STATe] on page 132
Querying spot noise results on 10
x
offset frequencies:
CALCulate<n>:SNOise:DECades:X? on page 132
CALCulate<n>:SNOise:DECades:Y? on page 132
On User Defined Offsets / Offset Frequency
Turns custom spot noise frequencies on and off.
If on, the "Offset Frequency" input fields become available. You can measure the spot
noise for up to five custom offset frequencies. If active, the application adds those
spots to the spot noise table.
Remote command:
Turning spot noise marker on and off:
CALCulate<n>:SNOise<m>:STATe on page 133
CALCulate<n>:SNOise:AOFF on page 131
Positioning spot noise markers:
CALCulate<n>:SNOise<m>:X on page 133
Querying custom spot noise results:
CALCulate<n>:SNOise<m>:Y? on page 134
6.3 Using Limit Lines
Access: "Overview" > "Limit Analysis"
The "Limits" tab of the "Results" dialog box contains all funtions necessary to set up
and configure limit lines.
Using Limit Lines