Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
132User Manual 1173.9286.02 ─ 15
Manual operation: See "On User Defined Offsets / Offset Frequency" on page 58
CALCulate<n>:SNOise:DECades[:STATe] <State>
This command turns the spot noise calculation on every 10
x
offset frequency on and
off.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example:
CALC:SNO:DEC ON
Turns the spot noise calculation for each decade start on.
Manual operation: See "On All Decade Edges" on page 58
CALCulate<n>:SNOise:DECades:X?
This command queries the horizontal poistion of the 10
x
offset frequency spot noise
markers.
Suffix:
<n>
.
Window
Return values:
<OffsetFrequency>
List of offset frequencies, one for each 10
x
spot noise marker.
The number of return values depends on the measurement
range.
Default unit: Hz
Example:
CALC:SNO:DEC:X?
Return values, e.g.:
1000,10000,100000,1000000
Usage: Query only
Manual operation: See "Spot Noise" on page 16
See "On All Decade Edges" on page 58
CALCulate<n>:SNOise:DECades:Y?
This command queries the vertical poistion of the 10
x
offset frequency spot noise mark-
ers.
Suffix:
<n>
.
Window
Configure Numerical Result Displays