Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
133User Manual 1173.9286.02 ─ 15
Return values:
<Level>
List of level values, one for each 10
x
spot noise marker. The
number of return values depends on the measurement range.
Default unit: dBc/Hz
Example:
CALC:SNO:DEC:Y?
Return values, e.g.:
-152.560974121094,-136.443389892578,
-145.932891845703,-152.560974121094
Usage: Query only
Manual operation: See "Spot Noise" on page 16
See "On All Decade Edges" on page 58
CALCulate<n>:SNOise<m>:STATe
<State>
This command turns a custom spot noise marker on and off.
Suffix:
<n>
.
Window
<m> 1...5
Selects the spot noise marker.
Parameters:
<State> ON | OFF | 1 | 0
*RST: All 1
Example:
CALC:SNO3:STATE ON
Turns spot noise marker 3 on.
Manual operation: See "On User Defined Offsets / Offset Frequency" on page 58
CALCulate<n>:SNOise<m>:X <OffsetFrequency>
This command defines the horizontal position of a custom spot noise marker.
Suffix:
<n>
.
Window
<m> 1...5
Selects the spot noise marker.
Parameters:
<OffsetFrequency> The minimum offset is 1 Hz. The maximum offset depends on
the hardware you are using.
The default value varies for each of the five spot noise markers.
For marker 1 it is 1 kHz, for marker 2 it is 10 kHz, for marker 3 it
is 100 kHz, for marker 4 it is 1 MHz and for marker 5 it is
10 MHz