Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
95User Manual 1173.9286.02 ─ 15
Parameters:
<Gain> 15 dB | 30 dB
The availability of gain levels depends on the model of the
R&S FSW.
R&S FSW8/13: 15dB and 30 dB
R&S FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST: OFF
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation: See " Preamplifier " on page 39
8.7 Controlling the Measurement
The following commands are necessary to control the sequence of the phase noise
measurement.
[SENSe:]FREQuency:TRACk............................................................................................95
[SENSe:]FREQuency:VERify:TOLerance:ABSolute.............................................................95
[SENSe:]FREQuency:VERify:TOLerance:RELative............................................................. 96
[SENSe:]FREQuency:VERify[:STATe]................................................................................96
[SENSe:]IQ:DECimation................................................................................................... 96
[SENSe:]IQ:DPLL.............................................................................................................97
[SENSe:]IQ:ONLine..........................................................................................................97
[SENSe:]IQ:TBW............................................................................................................. 97
[SENSe:]POWer:RLEVel:VERify:TOLerance.......................................................................98
[SENSe:]POWer:RLEVel:VERify[:STATe]...........................................................................98
[SENSe:]POWer:TRACk...................................................................................................98
[SENSe:]REJect:AM.........................................................................................................98
[SENSe:]SWEep:SVFailed................................................................................................99
[SENSe:]FREQuency:TRACk <State>
This command turns frequency tracking on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example:
FREQ:TRAC OFF
Turns off frequency tracking.
Manual operation: See "Frequency Tracking" on page 41
[SENSe:]FREQuency:VERify:TOLerance:ABSolute <Frequency>
This command defines an absolute frequency tolerance for frequency verification.
Controlling the Measurement