Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
97User Manual 1173.9286.02 ─ 15
Example:
IQ:DEC ON
Turns on decimation.
Manual operation: See "Decimation" on page 42
[SENSe:]IQ:DPLL <State>
This command turns the digital PLL on and off.
Parameters:
<State> ON | OFF | 1 | 0
Example:
IQ:DPLL ON
Turns the digital PLL on.
Manual operation: See "Digital PLL" on page 42
[SENSe:]IQ:ONLine <State>
This command turns the I/Q online measurement mode on and off.
This mode is available for offset frequencies smaller than 30 kHz.
Note that you have to
turn on decimation with [SENSe:]IQ:DECimation
select the I/Q FFT mode for the affected half decades with [SENSe:
]BANDwidth|BWIDth[:RESolution]:TYPE
turn off forward sweep with [SENSe:]SWEep:FORWard
Parameters:
<State> ON | OFF | 1 | 0
Example:
SWE:FORW OFF
LIST:RANG9:FILT:TYPE IQFF
LIST:RANG8:FILT:TYPE IQFF
IQ:DEC ON
IQ:ONL ON
Configures a online measurement for the half decades between
3 kHz and 10 kHz and 10 kHz and 30 kHz.
Manual operation: See "Online I/Q" on page 42
[SENSe:]IQ:TBW <Bandwidth>
This command defines the maximum tracking bandwidth (sample rate) for all half deca-
des.
Parameters:
<Bandwidth> Range: 60 mHz to 65.28 MHz
Increment: 10 mHz
*RST: 60 mHz