EasyManua.ls Logo

Rohde & Schwarz FSW-K40 - Page 100

Rohde & Schwarz FSW-K40
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
100User Manual 1173.9286.02 ─ 15
Suffix Half Decade
17 100 MHz ... 300 MHz
18 300 MHz ... 1 GHz
19 1 GHz ... 3 GHz
20 3 GHz ... 10 GHz
[SENSe:]LIST:BWIDth[:RESolution]:RATio....................................................................... 100
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE.............................................................. 100
[SENSe:]FREQuency:STARt...........................................................................................101
[SENSe:]FREQuency:STOP............................................................................................101
[SENSe:]LIST:BWIDth:RESolution:TYPE..........................................................................101
[SENSe]:LIST:IQWindow:TYPE.......................................................................................102
[SENSe:]LIST:RANGe<range>:BWIDth[:RESolution].........................................................102
[SENSe:]LIST:RANGe<range>:FILTer:TYPE.................................................................... 103
[SENSe:]LIST:RANGe<range>:IQWindow:TYPE...............................................................103
[SENSe:]LIST:RANGe<range>:SWEep:COUNt.................................................................104
[SENSe:]LIST:SWEep:COUNt.........................................................................................104
[SENSe:]LIST:SWEep:COUNt:MULTiplier........................................................................ 104
[SENSe:]LIST:SWEep:COUNt:MULTiplier:STATe............................................................. 105
[SENSe:]SWEep:FORWard............................................................................................ 105
[SENSe:]SWEep:MODE................................................................................................. 105
[SENSe:]LIST:BWIDth[:RESolution]:RATio <Ratio>
This command defines the resolution bandwidth over all half decades.
Parameters:
<Ratio> Numeric value in %.
The resulting RBW is the percentage of the start frequency of
each half decade.
If the resulting RBW is not available, the application rounds to
the next available bandwidth.
Range: 1 to 100
*RST: 10
Example:
LIST:BWID:RAT 20
Defines a RBW of 20% of the start frequency of the correspond-
ing half decade.
Manual operation: See "Global RBW" on page 44
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE <Mode>
This command selects the sweep mode for a all half decades globally.
Configuring the Measurement Range

Table of Contents

Related product manuals