Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
102User Manual 1173.9286.02 ─ 15
Parameters:
<Mode> IQFFt
Measurement based on I/Q data.
NORMal
Measurement based on spectrum analyzer data.
FFT
Measurement based on spectrum analyzer data. Kept for com-
patibility to R&S FSV.
*RST: Depends on half decade
Example:
LIST:BWID:RES:TYPE IQFF
Selects I/Q analysis mode for all half decades.
Manual operation: See "Global Sweep Mode" on page 45
[SENSe]:LIST:IQWindow:TYPE
<WindowFunction>
This command selects the window function for all half decades.
Window functions are available for I/Q sweep mode.
Parameters:
<WindowFunction> RECtangular
GAUSsian
CHEBychev
BHARris
*RST: Depends on the half decade
Example:
LIST:IQW:TYPE REC
Selects a rectangular FFT window.
Manual operation: See "Global I/Q Window" on page 45
[SENSe:]LIST:RANGe<range>:BWIDth[:RESolution] <RBW>
This command defines the resolution bandwidth for a particular half decade.
Suffix:
<range>
.
1...20
Selects the half decade.
For the suffix assignment see Table 8-4.
Parameters:
<RBW> Numeric value in Hz.
Note that each half decade has a limited range of available
bandwidths.
*RST: Depends on the half decade
Default unit: Hz
Configuring the Measurement Range