Configuration
R&S
®
FSW-K40
45User Manual 1173.9286.02 ─ 15
You can also change the global bandwidth with the "RBW Global" softkey in the "Band-
width" menu.
Remote command:
[SENSe:]LIST:BWIDth[:RESolution]:RATio on page 100
Global Average Count
Defines the number of measurements that the application uses to calculate averaged
results in each half decade.
The range is 1 to 10000.
Remote command:
[SENSe:]LIST:SWEep:COUNt on page 104
Multiplier
Turns a multiplier that changes the average count in each half decade on and off.
If on, you can define a value that multiplies the number of averages currently defined
for each half decade by that value.
When you turn it off, the original averages are restored and used again.
Example:
You have three half decades:
●
1st half decade average count: 1
●
2nd half decade average count: 3
●
3rd half decade average count: 5
If you turn the multiplier on and define a value of 5, the average count changes as fol-
lows:
●
1st half decade average count: 5
●
2nd half decade average count: 15
●
3rd half decade average count: 25
Remote command:
[SENSe:]LIST:SWEep:COUNt:MULTiplier on page 104
[SENSe:]LIST:SWEep:COUNt:MULTiplier:STATe on page 105
Global Sweep Mode
Selects the analysis mode for all half decades. The sweep mode defines the way the
application processes the data.
For more information see Chapter 4.4, "Sweep Modes", on page 25.
"Normal"
Uses spectrum analyzer data for the data analysis.
"I/Q / FFT"
Uses I/Q data for the data analysis.
Remote command:
[SENSe:]LIST:BWIDth:RESolution:TYPE on page 101
Global I/Q Window
Selects the window function for all half decades.
The window function is available for I/Q analysis.
Configuring the Measurement Range