Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
105User Manual 1173.9286.02 ─ 15
[SENSe:]LIST:SWEep:COUNt:MULTiplier:STATe <State>
This command turns a multiplier that is applied to the average count in each half dec-
ade on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example: See [SENSe:]LIST:SWEep:COUNt:MULTiplier
on page 104.
Manual operation: See "Multiplier" on page 45
[SENSe:]SWEep:FORWard <State>
This command selects the measurement direction.
Specifies the sweep direction. When switched on the sweep direction is from the start
frequency to the stop frequency. When switched off the sweep direction is reversed
Parameters:
<State> ON | 1
Measurements in forward direction.
The measurements starts at the smallest offset frequency.
OFF | 0
Measurement in reverse direction.
The measurement starts at the highest offset frequency.
*RST: 0
Example:
SWE:FORW ON
Selects forward measurements.
Manual operation: See "Sweep Forward" on page 44
[SENSe:]SWEep:MODE <Mode>
This command selects the type of measurement configuration.
Parameters:
<Mode> AVERage
Selects a measurement configuration optimized for quality
results.
FAST
Selects a measurement configuration optimized for speed.
MANual
Selects manual measurement configuration.
NORMal
Selects a balanced measurement configuration.
*RST: NORMal
Manual operation: See "Presets" on page 44
Configuring the Measurement Range