Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
104User Manual 1173.9286.02 ─ 15
Manual operation: See "Half Decades Configuration Table" on page 46
[SENSe:]LIST:RANGe<range>:SWEep:COUNt <Measurements>
This command defines the number of measurements included in the averaging for a
half decade.
Suffix:
<range>
.
1...20
Selects the half decade.
For the suffix assignment see Table 8-4.
Parameters:
<Measurements>
Range: 1 to 10000
*RST: Depends on the half decade
Example:
LIST:RANG9:SWE:COUN 15
Includes 15 measurements in the averaging of the ninth half
decade.
Manual operation: See "Half Decades Configuration Table" on page 46
[SENSe:]LIST:SWEep:COUNt <Averages>
This command defines the number of measurements to be included in the averaging
for each and all half decades.
Parameters:
<Averages> Range: 1 to 10000
*RST: Depends on the half decade.
Example:
LIST:SWE:COUN 20
Defines 20 averages for all half decades.
Manual operation: See "Global Average Count" on page 45
[SENSe:]LIST:SWEep:COUNt:MULTiplier <Multiplier>
This command defines a multiplier that is applied to the average count in each half
decade.
Before you can use the command you have to turn on the multiplier with [SENSe:
]LIST:SWEep:COUNt:MULTiplier:STATe.
Parameters:
<Multiplier> Numeric value.
Example:
LIST:SWE:COUN:MULT:STAT ON
LIST:SWE:COUN:MULT 5
Turns on the multiplier and multiplies the averages by 5.
Manual operation: See "Multiplier" on page 45