Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
103User Manual 1173.9286.02 ─ 15
Example:
LIST:RANG9:BWID 100Hz
Selects a RBW of 100 Hz for the half decade from 1 kHz to 3
kHz.
Manual operation: See "Half Decades Configuration Table" on page 46
[SENSe:]LIST:RANGe<range>:FILTer:TYPE <Mode>
This command selects the sweep mode for a particular half decade.
Suffix:
<range>
.
1...20
Selects the half decade.
For the suffix assignment see Table 8-4.
Parameters:
<Mode> IQFFt
Measurement based on I/Q data.
NORMal
Measurement based on spectrum analyzer data.
FFT
Measurement based on spectrum analyzer data. Kept for com-
patibility to R&S FSV.
*RST: Depends on half decade
Example:
LIST:RANG9:FILT:TYPE FFT
Selects FFT analysis for the ninth half decade.
Manual operation: See "Half Decades Configuration Table" on page 46
[SENSe:]LIST:RANGe<range>:IQWindow:TYPE <WindowFunction>
This command selects the window function for a particular half decade.
Window functions are available for I/Q sweep mode.
Suffix:
<range>
.
1...20
Selects the half decade.
For the suffix assignment see Table 8-4.
Parameters:
<WindowFunction> RECtangular
GAUSsian
CHEBychev
BHARris
*RST: Depends on the half decade
Example:
LIST:RANG:IQW:TYPE BHAR
Selects the Blackman Harris window function for the first half
decade.
Configuring the Measurement Range