Loading...
Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
144User Manual 1173.9286.02 ─ 15
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage: Query only
8.12.3 Configuring Markers
DISPlay:MINFo[:STAT]...................................................................................................144
DISPlay:MTABle............................................................................................................ 144
DISPlay:MINFo[:STAT] <DisplayMode>
This command turns the marker information in all diagrams on and off.
Parameters:
<DisplayMode> ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST: 1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation: See " Marker Info " on page 67
DISPlay:MTABle <DisplayMode>
This command turns the marker table on and off.
Parameters:
<DisplayMode> ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST: AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation: See " Marker Table Display " on page 67