Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
85User Manual 1173.9286.02 ─ 15
SYSTem:SEQuencer <State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (INIT:SEQ...) are executed, otherwise an error will
occur.
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S FSW User Manual.
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands (INIT:
SEQ...) are not available.
*RST: 0
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF
8.5 Configuring the Result Display
The commands required to configure the screen display in a remote environment are
described here.
● General Window Commands.................................................................................. 85
● Working with Windows in the Display..................................................................... 86
8.5.1 General Window Commands
The following commands are required to configure general window layout, independent
of the application.
DISPlay:FORMat............................................................................................................. 85
DISPlay[:WINDow<n>]:SIZE............................................................................................. 86
DISPlay:FORMat <Format>
This command determines which tab is displayed.
Configuring the Result Display