Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
101User Manual 1173.9286.02 ─ 15
Parameters:
<Mode> IQ
Measurement based on I/Q data.
NORMal
Measurement based on spectrum analyzer data.
FFT
Measurement based on spectrum analyzer data. Kept for com-
patibility to R&S FSV.
*RST: Depends on half decade
Example:
BAND:TYPE FFT
Selects FFT analysis for all half decades.
[SENSe:]FREQuency:STARt <Frequency>
This command defines the start frequency of the measurement range.
Parameters:
<Frequency> Offset frequencies in half decade steps.
Range: 1 Hz to 3 GHz
*RST: 1 kHz
Example:
FREQ:STAR 10kHz
Defines a start frequency of 10 kHz.
Manual operation: See "Range Start / Stop" on page 44
[SENSe:]FREQuency:STOP <Frequency>
This command defines the stop frequency of the measurement range.
Parameters:
<Frequency> Offset frequencies in half decade steps.
Range: 3 Hz to 10 GHz
*RST: 1 MHz
Example:
FREQ:STOP 10MHz
Defines a stop frequency of 10 MHz.
Manual operation: See "Range Start / Stop" on page 44
[SENSe:]LIST:BWIDth:RESolution:TYPE <Mode>
This command selects the sweep mode for all half decades.
Configuring the Measurement Range