Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
96User Manual 1173.9286.02 ─ 15
If you define both an absolute and relative tolerance, the application uses the higher
tolerance level.
Parameters:
<Frequency> Numeric value in Hz.
*RST: 1 kHz
Example:
FREQ:VER:TOL:ABS 100kHz
Defines a frequency tolerance range of 100 kHz.
Manual operation: See "Verify Frequency" on page 40
[SENSe:]FREQuency:VERify:TOLerance:RELative <Percentage>
This command defines a relative frequency tolerance for frequency verification.
If you define both an absolute and relative tolerance, the application uses the higher
tolerance level.
Parameters:
<Percentage> Numeric value in %, relative to the current nominal frequency.
Range: 1 to 100
*RST: 10
Default unit: PCT
Example:
FREQ:VER:TOL:REL 12
Defines a frequency tolerance of 12% in relation to the nominal
frequency.
Manual operation: See "Verify Frequency" on page 40
[SENSe:]FREQuency:VERify[:STATe] <State>
This command turns frequency verification on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example:
FREQ:VER ON
Turns on frequency verification.
Manual operation: See "Verify Frequency" on page 40
[SENSe:]IQ:DECimation <State>
This command turns decimation of results on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Controlling the Measurement