Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
98User Manual 1173.9286.02 ─ 15
Example:
IQ:TBW 100HZ
Defines a tracking bandwidth of 100 Hz.
Manual operation: See "Max Freq Drift" on page 42
[SENSe:]POWer:RLEVel:VERify:TOLerance <Level>
This command defines a relative level tolerance for level verification
Parameters:
<Level> Numeric value in dB, relative to the nominal level.
*RST: 10 dB
Example:
POW:RLEV:TOL 5
Defines a level tolerance of 5 dB.
Manual operation: See "Verify Level" on page 41
[SENSe:]POWer:RLEVel:VERify[:STATe] <State>
This command turns level verification on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example:
POW:RLEV:VER ON
Turns on level verification.
Manual operation: See "Verify Level" on page 41
[SENSe:]POWer:TRACk <State>
This command turns level tracking on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example:
POW:TRAC ON
Turns on level tracking.
Manual operation: See "Level Tracking" on page 41
[SENSe:]REJect:AM <State>
This command turns the suppression of AM noise on and off.
Parameters:
<State> ON | OFF | 1 | 0
Example:
REJ:AM ON
Turns AM noise suppression on.