Configuration
R&S
®
FSW-K40
41User Manual 1173.9286.02 ─ 15
Absolute tolerance:
[SENSe:]FREQuency:VERify:TOLerance:ABSolute on page 95
Verify Level
Turns level verification on and off.
If level verification is on, the R&S FSW initiates the phase noise measurement only if
the level of the DUT is within a certain level tolerance range. The tolerance range is a
level range relative to the nominal level.
For more information see Chapter 4.7, "Level Determination", on page 30.
Remote command:
Verify level:
[SENSe:]POWer:RLEVel:VERify[:STATe] on page 98
Level tolerance:
[SENSe:]POWer:RLEVel:VERify:TOLerance on page 98
On Verify Failed
Selects the way the application reacts if signal verification fails.
Takes effect on both frequency and level verification.
"Restart"
Restarts the measurement if verification has failed.
"Stop"
Stops the measurement if verification has failed.
"Run Auto All"
Starts an automatic frequency and level detection routine if verifica-
tion has failed. After the new frequency and level have been set, the
measurement restarts. For more information see Chapter 5.8, "Auto-
matic Measurement Configuration", on page 49.
Remote command:
[SENSe:]SWEep:SVFailed on page 99
Frequency Tracking
Turns frequency tracking on and off.
If on, the application tracks the frequency of the DUT during the phase noise measure-
ment and adjusts the nominal frequency accordingly. The application adjusts the fre-
quency after each half decade measurement.
For more information see Chapter 4.6, "Frequency Determination", on page 27.
Remote command:
[SENSe:]FREQuency:TRACk on page 95
Level Tracking
Turns level tracking on and off.
If on, the R&S FSW tracks the level of the DUT during phase noise measurements and
adjusts the nominal level accordingly. The application adjusts the level after each half
decade measurement.
For more information see Chapter 4.7, "Level Determination", on page 30.
Remote command:
[SENSe:]POWer:TRACk on page 98
Controlling the Measurement