Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
125User Manual 1173.9286.02 ─ 15
Manual operation: See "Spur Removal / Spur Threshold" on page 56
[SENSe:]SPURs:THReshold <Threshold>
This command defines the level threshold for spur removal.
Parameters:
<Threshold> Range: 0 to 50
*RST: 0
Default unit: dB
Example:
SPUR:THR 10
Defines a spur threshold of 50 dB.
Manual operation: See "Spur Removal / Spur Threshold" on page 56
TRACe<n>:COPY
<TraceNumber>, <TraceNumber>
This command copies data from one trace to another.
Suffix:
<n>
.
Window
Parameters:
<TraceNumber>,
<TraceNumber>
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
The first parameter is the destination trace, the second parame-
ter is the source.
(Note the 'e' in the parameter is required!)
Example:
TRAC:COPY TRACE1,TRACE2
Copies the data from trace 2 to trace 1.
Manual operation: See " Copy Trace " on page 55
TRACe[:DATA]? <Trace>
This command queries the results of the graphical result displays.
Parameters:
<Trace> TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
Return values:
<Frequency>,
<Level>
Phase noise:
Coordinates of the phase noise trace as list of comma separated
values, beginning at the nearest offset frequency.
<OffsetFrequency1>,<Level1>,<OffsetFrequency2>,<Level2>,...
Spectrum monitor:
Coordinates of the spectrum trace as a list of comma separated
values, beginning at the left border of the display.
<Frequency1>,<Level1>,<Frequency2>,<Level2>,...
Graphical Display of Phase Noise Results