EasyManua.ls Logo

Rohde & Schwarz FSW-K40 - Page 130

Rohde & Schwarz FSW-K40
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
130User Manual 1173.9286.02 ─ 15
Suffix:
<t>
.
Trace
Example:
FETC:PNO:RPM?
Queries the residual PM for the first trace of the diagram.
Usage: Query only
Manual operation: See "Residual Noise" on page 15
FETCh:PNOise<t>:USER<range>:IPN?
This command queries the Integrated Phase Noise for a particular user range.
The trace that is queried depends on CALCulate<n>:EVALuation:USER<range>:
TRACe.
Suffix:
<t>
.
Trace
<range> 1...3
Selects the user range.
Example:
FETC:PNO:USER2:IPN?
Queries the Integrated Phase Noise for user range 2.
Usage: Query only
FETCh:PNOise<t>:USER<range>:RFM?
This command queries the residual FM for a particular user range.
The trace that is queried depends on CALCulate<n>:EVALuation:USER<range>:
TRACe.
Suffix:
<t>
.
Trace
<range> 1...3
Selects the user range.
Example:
FETC:PNO:USER2:RFM?
Queries the residual FM for user range 2.
Usage: Query only
Manual operation: See "Residual Noise" on page 15
FETCh:PNOise<t>:USER<range>:RMS?
This command queries the residual RMS jitter for a particular user range.
The trace that is queried depends on CALCulate<n>:EVALuation:USER<range>:
TRACe.